PASCO Specialty & Mfg. CI-6734 Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Medição, teste e controlo PASCO Specialty & Mfg. CI-6734. PASCO Specialty & Mfg. CI-6734 User's Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
better
teach science
ways to
Phone (916) 786-3800 • FAX (916) 786-8905 • web: www.pasco.com
10101 Foothills Blvd. • P.O. Box 619011 • Roseville, CA 95678-9011 USA
®
© 1997 PASCO scientific $7.50
Instruction Manual and
Experiment Guide for the
PASCO scientific
Model CI-6734
012-06615A
9/97
SODIUM ION SELECTIVE
ELECTRODE
Vista de página 0
1 2 3 4 5 6 ... 21 22

Resumo do Conteúdo

Página 1 - ELECTRODE

betterteach scienceways toPhone (916) 786-3800 • FAX (916) 786-8905 • web: www.pasco.com10101 Foothills Blvd. • P.O. Box 619011 • Roseville, CA 95678-

Página 2 - Always use eye protection

Sodium Ion Selective Electrode 012–06615A6Measurement ProcedureDirect MeasurementDirect measurement is a simple procedure for measuring a large number

Página 3 - Table of Contents

012–06615A Sodium Ion Selective Electrode7Figure 3Typical sodium electrode calibration curve6. To a clean, dry 150 ml beaker, add 100 ml of sample and

Página 4

Sodium Ion Selective Electrode 012–06615A86. Add increments of the l.0xl0-3 M or 100 ppm standard as given in Table 2 below.7. After the reading has s

Página 5 - Introduction

012–06615A Sodium Ion Selective Electrode9Electrode drift and slow response could indicate the presence of high interference from the ions listed. Soa

Página 6 - Equipment

Sodium Ion Selective Electrode 012–06615A10Electrode ResponsePlotting the electrode potential against the sodium concentration on semilogarithmic pape

Página 7

012–06615A Sodium Ion Selective Electrode11Figure 5Electrode potential behavior vs. solution pH in pure NaCl solution at 25 °The edge of the shaded ar

Página 8 - General Preparation

Sodium Ion Selective Electrode 012–06615A12SpecificationsConcentration Range: Saturated to 1.0Xl0-6 M (0.02 ppm)pH Range: 5 to 12 (depending on pH lev

Página 9 - Measurement

012–06615A Sodium Ion Selective Electrode13Standards & ReagentsWhenever problems arise with the measuring procedure that has been used successfull

Página 10 - Measurement Procedure

Sodium Ion Selective Electrode 012–06615A14Symptom Possible Causes Next StepNoisy or Unstable electrode exposed to interferences soak overnight in ele

Página 11

012–06615A Sodium Ion Selective Electrode15Symptom Possible Causes Next Step“Incorrect Answer” incorrect scaling of semilog paper plot voltage potenti

Página 12 - Electrode Characteristics

Sodium Ion Selective Electrode 012–06615AAlways use eye protectionand gloves when workingwith chemicals.

Página 13 - Temperature Influences

Sodium Ion Selective Electrode 012–06615A16

Página 14 - Limits of Detection

Technical SupportFeedbackIf you have any comments about the product ormanual, please let us know. If you have anysuggestions on alternate experiments

Página 16 - Specifications

012–06615A Sodium Ion Selective ElectrodeTable of ContentsIntroduction...

Página 17 - Troubleshooting Hints

iiEquipment ReturnShould the product have to be returned to PASCOscientific for any reason, notify PASCO scientific byletter, phone, or fax BEFORE ret

Página 18

012–06615A Sodium Ion Selective Electrode1IntroductionThe PASCO scientific Sodium Ion Selective Electrode is used to quickly, simply, accurately, and

Página 19

Sodium Ion Selective Electrode 012–06615A2ROO1013SAMPLE4MKCIReferenceFillSolutionFigure 1.Included Equipmentto ISEAmplifierSodium Ion SelectiveElectro

Página 20

012–06615A Sodium Ion Selective Electrode3 Note: Electrodes must not be stored in distilled water or air. Do not rinse withdistilled water.• Dilute E

Página 21 - Technical Support

Sodium Ion Selective Electrode 012–06615A4General PreparationElectrode Preparation1. Remove the rubber cap covering the electrode tip andthe rubber in

Página 22

012–06615A Sodium Ion Selective Electrode5MeasurementMeasuring Hints• All samples and standards should be at the same temperature for precise measurem

Comentários a estes Manuais

Sem comentários